Gel Filtration Chromatography
SW
SWxl
TSKgel G2000SWxl
TSKgel G3000SWxl
TSKgel G4000SWxl
SuperSW
Desalting Columns


TSKgel SWxl columns, introduced in 1987, represent the second generation of high performance Gel Filtration columns that have become synonymous with analyzing protein molecular weights in the emerging field of biotechnology.

Particle Characteristics
TSKgel SWxl-type columns contain smaller particles than TSKgel SW-type columns; 5 and 8μm versus 10 and 13μm. As the TSKgel SW-type columns, TSKgel SWxl columns feature highly porous silica particles, the surface of which has been shielded from interacting with proteins by derivatization with ligands containing diol functional groups. TSKgel SWxl-type columns stand out from other silica- or polymer-based high performance size exclusion columns by virtue of their large pore volumes. These columns are commonly used in the quality control of monoclonal antibodies and other biopharma products.

Calibration Curves
The Figure below shows the calibration curves for globular proteins, polyethylene oxides and dextrans for each of the three TSKgel SWxl-type columns. The following Table summarizes the molecular weight ranges for TSKgel G2000SWxl, G3000SWxl and G4000SWxl columns for globular proteins, polyethylene oxides and dextrans that can, at least partially, access the pores. For comparative purposes, similar data is provided as well for TSKgel SW and SuperSW-type columns.

Protein Calibration curves for
TSKgel SWxl columns





Properties and separation ranges for TSKgel SW-type packings (Molecular Weight in Da)

TSKgel   packing

Particle
size (µm)

Pore
size (Å)

Globular
proteins

Dextrans

Polyethylene
glycols and oxides

SuperSW2000

4

125

5,000 - 1.5 x 10^5

1,000 - 3 x 10^4

500 - 15,000

G2000SWxl
BioAssist G2SWxl

5

125

5,000 - 1.5 x 10^5

1,000 - 3 x 10^4

500 - 15,000

QC-PAK TSK200

5

125

5,000 - 1.5 x 10^5

1,000 - 3 x 10^4

500 - 15,000

G2000SW

10, 13

125

5,000 - 1 x 10^5

1,000 - 3 x 10^4

500 - 15,000

SuperSW3000

4

250

1 x 10^4 - 5 x 10^5

2,000 - 7 x 10^4

1,000 - 3.5 x 10^4

G3000SWxl
BioAssist G3SWxl

5

250

1 x 10^4 - 5 x 10^5

2,000 - 7 x 10^4

1,000 - 3.5 x 10^4

QC-PAK TSK300

5

250

1 x 10^4 - 5 x 10^5

2,000 - 7 x 10^4

1,000 - 3.5 x 10^4

G3000SW

10

250

1 x 10^4 - 5 x 10^5

2,000 - 7 x 10^4

1,000 - 3.5 x 10^4

           

G4000SWxl
BioAssist G4SWxl

8

450

2 x 10^4 - 7 x 10^6

4,000 - 5 x 10^5

2,000 - 2.5 x 10^5

G4000SW

13, 17

450

2 x 10^4 - 7 x 10^6

4,000 - 5 x 10^5

2,000 - 2.5 x 10^5

Data generated using the following conditions:

Columns:

Two 4µm, 4.6mm ID x 30cm TSKgel SuperSW columns in series; two 5µm, 7.8mm ID x 30cm TSKgel SWxl columns in series; two 10µm, 7.5mm ID x 60cm TSKgelSW columns in series

Elution:

Globular proteins: 0.3M NaCl in 0.1M (0.05M for TSKgel SWxl columns) phosphate buffer, pH 7
Dextrans and polyethylene glycols and oxides (PEOs): distilled water

 


Contact us:

Tech Service:
800-366-4875 (option #3)

Customer Service:
866-527-3587

info.tbl@tosoh.com
 


Return to HPLC Columns main page

Return to Size Exclusion main page

Application Overview -- Gel Filtration Chromatography with TSKgel columns

Tosoh Bioscience LLC 3604 Horizon Drive, Suite 100, King of Prussia, PA, 19406 Tel: (800) 366-4875  Email: info.tbl@tosoh.com