Questions:

How do I access Tosoh’s HPLC Application Database?

What is the cause of negative peaks?

How do I assemble a stainless steel guard column kit?

Is it always necessary to use a guard column with a TSKgel analytical column?

What is the function of a guard column? What makes a good guard column? What are the different guard column designs? How do I install a guard column?

How can you tell when it is time to change your guard column?

What endfitting do I use with my particular TSKgel column?

What does TSKgel stand for?

What is the significance of “-5PW” when referring to TSKgel columns?

What kind of fittings and connectors will I need in order to use TSKgel HPLC columns?

What quality control procedures are used by Tosoh Bioscience to assure column performance?

Please explain the lot numbering system used for TSKgel columns.


1. How do I access Tosoh’s HPLC Application Database? (back to top)

The Tosoh Bioscience web site contains a direct link to the Tosoh Corporation HPLC database. This database contains many application examples that demonstrate the performance of Tosoh’s chromatographic products. The sources of these applications include published articles in scientific journals as well as technical papers and presentations generated by Tosoh scientists.

To access the database, simply click here or return to the Resource Center homepage and use the links listed on that page to navigate to the database.  

There are three main categories to aid in your search:

  • ANY DATA is where you can search all information registered in this database.  This database consists of 'Application Data' and 'Published Literature'.
  • APPLICATION DATA is specific for chromatographic information extracted from Tosoh’s original data or from customer’s submitted applications.
  • PUBLISHED LITERATURE is specific for information published in scientific journals that refer directly to Tosoh’s products.

2. What is the cause of negative peaks? (back to top)

A negative peak is a peak that dips below an established flat baseline. Negative peaks arise when the refractive index or UV absorbance of the solute is less than that of the mobile phase. When using an RI detector, simply reversing the polarity of the recording device will make the peak show up as a positive signal. The same ‘trick’ is applicable when using a UV detector, although this is rarely needed since most mobile phases have much lower UV absorptivity at the wavelength where the solutes have significant absorption.

Negative peaks are most often seen at the beginning of the chromatogram, where differences in the composition of the sample solvent and the mobile phase may give rise to an unretained solvent peak that first shows a positive deflection, but then goes negative before returning to the baseline. Since measurements are best performed when peaks elute after two or three column volumes have passed through the column, i.e., k=1 or 2, such ‘solvent’ peaks are usually ignored.

3. How do I assemble a stainless steel guard column kit? (back to top)

If you would rather watch a video on YouTube of how to assemble a guard column, please follow this link

TSKgel guard columns are shipped partially assembled. The first step in assembling a guard column is to set the ferrules onto the column. Place the column assembly on the table with the column facing up, as shown in the figure on the left below. Using wrenches, tighten the cap onto the reducing union by making approximately 1/2 turn. This should set the ferrules onto the column tube. Separate the cap and the column from the reducing union. Take the frit out of the reducing union and place both on the bench.

  1. Fit the cap and the column into a laboratory clamp.
  2. Attach tubing from an aspirator or other gentle vacuum source to the   column and apply suction.
  3. Resuspend the Guardgel in its container by adding about 10mL of the appropriate shipping solvent.
  4. Pipet the slurry into the column using a medicine dropper or pipet.
  5. Continue adding gel until the surface of the gel bed is flush with the top of the column.
  6. Remove excess packing with a spatula. Make sure that no packing is left on the ferrules.
  7. Disconnect the vacuum source.
  8. Place the frit and the reducing union on top of the packed guard column. Tighten with a wrench.
  9. Connect the guard column (at the reducing union side!) to the injector of your HPLC or LC system following the procedure described for the separation column in the Instruction Manual.
  10. Start the pump at a low flow rate (0.5mL/min) and slowly increase to 1mL/min.
  11. Connect the guard column to the analytical separation column after sufficient solvent has exited to expel all air from the guard column.

Guardgel  Kit Assembly

4. Is it always necessary to use a guard column with a TSK- GEL analytical column? (back to top)

A guard column is generally recommended because it protects the analytical column from a major contamination of the HPLC, e.g. pump seal failure.  It also protects against inadvertent overloading of the column because a guard column is both a particulate AND molecular filter.  However, if the mobile phase and sample are filtered to 0.22- 0.45 micron a guard column may not be necessary.  The user still, however, exposes the column to the potential contamination from the HPLC system.

5. What is the function of a guard column? What makes a good guard column? What are the different guard column designs? How do I install a guard column? (back to top)

The guard column is installed between the injector and the analytical column and is designed to increase the lifetime of an analytical column by protecting the analytical column from unwanted materials, such as (1) highly retained or (2) irreversibly retained compounds and (3) particulate matter.

  1. Examples of highly retained compounds are fatty acids or oils when operating under reversed phase conditions (e.g., C18 column and methanol/water or acetonitrile/water mobile phase) where the retention factors of the fatty acids are infinite at the strongest mobile phase condition.
  2. Proteins can be examples of irreversibly retained compounds (in reversed phase HPLC) when operating under conditions in which proteins are not soluble in the mobile phase. This can happen when not all protein has been removed during an extraction before the sample is injected.
  3. Particulate matter may be introduced by the sample, if not appropriately filtered before injection. Particulates may also be shed by system components such as seals in the pump or injector.

What makes a good guard column?

  1. Since the role of a guard column is to prevent fouling of the analytical column, the guard column should be considerably less expensive than the analytical column it protects.
  2. To simplify method development the guard column should contain the same packing material as the analytical column.    

    a. A larger particle size is acceptable as long as the packing material has the same chemical characteristics as the packing in the analytical column.
    b. The use of guard columns filled with large particle size pellicular packings used to be popular. Pellicular guard columns are relatively cheap and are effective in protecting the analytical column from particulate matter, but their low retentive capacity made them less suitable to protect the column from highly retentive sample components. 
  3. To minimize the backpressure in the system, the length of the guard column should be as short as possible, while being long enough not to cause breakthrough of strongly retained solutes to the analytical column.
  4. To minimize the backpressure in the system, the diameter of the guard column should be the same as the ID of the analytical column.

    a. Since the flow rate of the analysis is determined by the requirements of the separation on the analytical column, a guard column with a smaller ID would cause extra back pressure over a guard column that has the same ID as the analytical column.
    b. This consideration is less important the shorter the length of the guard column (1-1.5cm). Since back pressure is proportional to ID2, a 1.5cm x 3.2mm ID guard column generates (4.6/3.2)2 ~ 2 times the back pressure of a 1.5cm x 4.6mm ID guard column.

  5. The top frit of the guard column has usually the same pore size as the frit on top of the analytical column, although the pore size can be larger if the guard column contains a larger particle size packing material. 
     
    a. The top frit should be small enough to prevent packing particles to block the frit when flow is momentarily interrupted during injection or when other sudden changes in back pressure occur.
    b. The top frit should be as large as possible to prevent frequent clogging from particulates derived from the sample or from system components.
    c. Since the top frit of the guard column is the most likely frit to get clogged during continued operation, it is helpful if the frit can be cleaned or replaced with relative ease.

What are the different guard column designs?
Guard columns are available either as small replaceable columns or as guard cartridges that are easily inserted into a hand-tight guard column holder.

How do I install a guard column?
Installing a guard column is performed the same way as a primary column is installed, using stainless steel or PEEK tubing and fittings.

6. How can you tell when it is time to change your guard column? (back to top)

Unfortunately guard columns are not like light bulbs that burn out when it is time to change them. It is best to get into a routine of changing guard columns on a regular basis like you would a water filter. Tosoh recommends replacing guard cartridges after every 30 to 40 injections although this frequency can vary depending on (1) the mode of chromatography, (2) the chemical nature of the sample, (3) the cleanliness of the sample in terms of particulate matter or the presence of highly retained components, the volume injected, etc. 

Signs that might indicate a need to change the guard column are higher than usual back pressure, peak broadening, changes in selectivity or changes in retention time. To determine the cause of the change in efficiency, selectivity, retention or pressure, it is recommended to remove the guard column and check the performance of the analytical column under the manufacturer’s QC conditions. Of course, one needs to always keep in mind that a malfunction of the HPLC system can also result in higher or lower than usual pressure, changes in retention and selectivity, and loss in column efficiency.

7. What endfitting do I use with my particular TSKgel column? (back to top)

Please visit the HPLC Columns - Replacement Parts page, which will provide a description of the available fittings, including the size of the frit in the fitting and the applicable endfitting to be used with your TSKgel column.  Question #4 below includes additional information on fittings and connectors.

8. What does TSKgel stand for? (back to top)

The original name of Tosoh Corporation, the parent company of Tosoh Bioscience, was Toyo Soda Manufacturing Co., Ltd.  TSK is short for Toyo Soda Kogyo (manufacturing in Japanese). The word  “gel” is the generic term for the resin used in our columns.

Tosoh owns a registered trademark for TSKgel. The mark TSKgel is used when we refer to a particular column, such as in a product description, and when referring to a product line or to multiple columns of the same type. TSKgel is also used as the brand name for the line of bulk resins in the 20 and 30 micron particle size range, which are used to pack high efficiency preparative columns in intermediate-scale manufacturing of biopharmaceutical drugs.


9. What is the significance of “-5PW” when referring to TSKgel columns? (back to top
)

Polymeric based Size Exclusion Columns supplied by Tosoh Bioscience start with a base support material.  One of these supports is methacrylate derivative with a pore size of 1000 Angstroms.  This is the same resin used in the TSKgel G5000PWxl columns.  So when this resin is functionalized with a ligand such as sulfo propyl the product is called SP (for sulfo propyl) -5PW (for the G5000PWxl support matrix).


10. What kind of fittings and connectors will I need in order to use TSKgel HPLC columns? (back to top)

The choice of fittings is dictated by the type of housing that contains the packing material. Most stainless steel and PEEK columns employ endfitings with 10-32 female inlet threads. Standard 1/16" OD stainless steel or PEEK tubing is used to connect the column to the guard column (or injector) and the detector. We recommend the use of “fingertight” style PEEK fittings with 1/16" OD PEEK tubing throughout your chromatography circuit.

Several of our polymer-based packings are also available in glass column format. These columns employ endfittings with ¼-28 female inlet threads. 
¼-28 fittings are typically used with 1/16” OD Teflon tubing for low to medium pressure applications.


11. What quality control procedures are used by Tosoh Bioscience to assure column performance? (back to top)

Tosoh Bioscience strives to attain the highest possible consistency and quality of performance for its columns and resins.  The support resin, e,g, silica, is tested thoroughly for both physical and chemical characteristics.  Once a ligand is attached to the base matrix the material is tested again according to stringent QC methods.  Finally, after the resin is packed the individual column is tested with an analyte.  This analysis is sent with the column on a sheet called “Inspection Data”.  EVERY TSKgel COLUMN IS TESTED INDIVIDUALLY BEFORE RELEASE FOR SALE AND THE INSPECTION DATA SHEET IS PROVIDED FOR FUTURE REFERENCE.


12. Please explain the lot numbering system used for TSKgel columns. (back to top)

The column numbering system consists of 2 parts:

  1. the column serial number which is unique to that specific column.
  2. the lot number which is the alphanumeric sequence after the dash in the column number.  For example: 

H4321-05H  Serial No: H4321  Lot No: 05H

The lot number refers to a production run of packing materials.  Occasionally resin lots have a small amount of material remaining in which case 2 or more of these small amounts are combined to make a “mixed” lot designated by the letter “M”, e.g.  -10HM.  

 


Technical Support

8:30AM - 5:00PM (EST)
Monday through Friday

US/Canada Toll Free
800-366-4875 (option #3)

techservice.tbl@tosoh.com


USP Reference for TSKgel columns


Return to Technical Service main page

Return to Resource Center main page

Return to FAQs main page

 

Tosoh Bioscience LLC 3604 Horizon Drive, Suite 100, King of Prussia, PA, 19406 Tel: (800) 366-4875  Email: info.tbl@tosoh.com